Contemporary Clinical Dentistry
   
  Home | About us | Editorial board | Search
Ahead of print | Current Issue | Archives | Advertise
Instructions | Online submission| Contact us | Subscribe |

 

Login  | Users Online: 5889  Print this pageEmail this pageSmall font sizeDefault font sizeIncrease font size 

ORIGINAL ARTICLE
Year : 2019  |  Volume : 10  |  Issue : 4  |  Page : 590-594

Comparative evaluation of three abutment–implant interfaces on stress distribution in and around different implant systems: A finite element analysis


1 Department of Prosthodontics, Nair Hospital Dental College, Mumbai, Maharashtra, India
2 Department of Prosthodontics, Government Dental College and Hospital, Mumbai, Maharashtra, India
3 Department of Pediatric and Preventive Dentistry, KLE's V.K. Institute of Dental Sciences, Belagavi, Karnataka, India

Correspondence Address:
Vaibhav Bandgar
No. 202, Prathamesh C.H.S.L., Panta Nagar Police Station, Ghatkopar East, Mumbai - 400 075, Maharashtra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ccd.ccd_739_18

Rights and Permissions

Aim: This study was conducted to compare and evaluate the effect of three abutment–implant connections on stress distribution around three different implants under similar material properties and loading condition using finite element analysis (FEA). Materials and Methods: Three different types of implant–abutment connections were selected. The features of these connections are Sample A: Tri-channel internal connection (Nobel Biocare); Sample B: Internal conical-hex Morse Taper (ADIN); and Sample C: Internal octa-Morse taper method (Osstem). The following softwares – ANSYS Version: 14.5 for FEA; Meshing software: Hypermesh 11; and CATIA: to produce computerized models of implants and for mandibular modeling were used. The implants were scanned with a high-quality scanner. All the above data were used to produce computerized models by CATIA software. Within the implant system, finite element method modeling was performed by implementing bonded conditions on the abutment–implant interfaces implementing four different load conditions. The computerized model was transferred to ANSYS software. A statistical analysis was done to compare the groups. Results: The samples were evaluated using three-dimensional FEA analysis. It was found that stress at 100 N, 100 N with 15° tilt, 300 N, and 300 N with 15° tilt was found to be highest in Sample A followed by Sample C and Sample B, and the difference was statistically insignificant. Conclusion: Within the limitations of the present study, the tri-channel internal connection showed maximum stresses and least by the internal conical-hex Morse Taper and internal octa-morse taper connection.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1757    
    Printed37    
    Emailed0    
    PDF Downloaded235    
    Comments [Add]    

Recommend this journal